| Controlled potential mode or potentiostat | |
| dc-potential range | ± 2 V |
| compliance voltage | ± 8 V |
| dc-potential resolution | 1 mV |
| dc-offset error | 2 mV |
| accuracy | ≤ 0.2% |
| ac-potential amplitude | 1 mV to 350 mV |
| current ranges | 1 nA to 10 mA (8 ranges) |
| maximum current | ± 10 mA |
| current resolution | 0.1% of current range |
| 1 pA on lowest current range | |
| accuracy | ≤ 0.3% of current range at 100 nA to 1 mA |
| ≤ 0.5% at 10 nA and ≤ 1% at 1 nA and 10 mA | |
| Controlled current mode (galvanostat) | |
| current ranges | 1 μA to 1 mA |
| dc-current range range | -2 to + 2 times selected current |
| dc-current resolution | 0.1% of selected current range |
| dc-offset error | ≤ 0.2% |
| current accuracy | ≤ 0.4% |
| maximum output voltage | ± 8 V |
| General: | |
| electrometer amplifier input | > 100 Gohm // 4 pF |
| rise time | approx. 50 μs |
| dimensions | 155 mm x 85 mm x 35 mm |
| temperature range | 0˚C to + 40˚C |
| weight | 0.42 kg |
| power | 2 AA cells NiMH for 6-8 hours operation, |
| battery charger included (6 V- 1500 mA) |