Hardware specifications: PalmSens instrument

Printer-friendly versionPrinter-friendly version
Productinfo Software specs Hardware specs Application note


Controlled potential mode or potentiostat  
dc-potential range ± 2 V
compliance voltage ± 8 V
dc-potential resolution 1 mV
dc-offset error 2 mV
accuracy ≤ 0.2%
ac-potential amplitude 1 mV to 350 mV
current ranges 1 nA to 10 mA (8 ranges)
maximum current ± 10 mA
   
current resolution 0.1% of current range
  1 pA on lowest current range
accuracy ≤ 0.3% of current range at 100 nA to 1 mA
  ≤ 0.5% at 10 nA and ≤ 1% at 1 nA and 10 mA
   
Controlled current mode (galvanostat)  
current ranges 1 μA to 1 mA
dc-current range range -2 to + 2 times selected current
dc-current resolution 0.1% of selected current range
dc-offset error ≤ 0.2%
current accuracy ≤ 0.4%
maximum output voltage ±  8 V
   
General:  
electrometer amplifier input > 100 Gohm // 4 pF
rise time approx. 50 μs
dimensions 155 mm x 85 mm x 35 mm
temperature range 0˚C to + 40˚C
weight 0.42 kg
power 2 AA cells NiMH for 6-8 hours operation,
  battery charger included (6 V- 1500 mA)

Specifications subject to change, Ivium Technologies © 2009