s-Module |
| System Performance: |
|
| 4 Electrodes |
WE, CE, RE and S |
| Current compliance |
±2.5 A (optional: ±5 A) |
| Maximum output Voltage |
±10 V below 1 A and ±8 V up to 2.5 A |
| Potentiostat Bandwidth |
250 kHz |
| Stability settings |
High Speed, Standard, and High Stability |
| Programmable response filter |
1 MHz , 100 kHz , 10 kHz , 1 kHz , 10 Hz |
| Signal acquisition |
dual channel 16 bit ADC, 100.000 samples/sec |
| |
|
| Potentiostat: |
|
| Applied potential range |
±10 V, with 0.33 mV resolution |
| Applied potential accuracy |
0.2%, or 2 mV |
| Current ranges |
±10 nA to ±10 A in 10 steps (optional: ±1pA to ±1nA) |
| Measured current resolution |
0.015% of current range, minimum 1 pA |
| |
|
| Galvanostat |
|
| Applied current resolution |
0.33% of applied current range |
| Applied current accuracy |
0.2% |
| Potential ranges |
±10mV, ±100mV, ±1V, ±10V |
| Measured potential resolution |
0.04% of potential range, minimum 400nV |
| |
|
| Impedance Analyser |
|
| Frequency range |
10µHz to 250kHz (optional: 10µHz to 1MHz) |
| Amplitude |
0.015mV to 1.0V, or 0.03% to 100% of current range |
| DC offset |
16 bit dc offset subtraction, and 2 dc-decoupling filters |
| |
|
| Electrometer |
|
| Input impedance |
>1000 Gohm //<8pF |
| Input bias current |
<10 pA |
| Bandwidth |
>1 MHz |
| |
|
| Peripheral connections |
|
| 2 Analog in |
0 to ±10 V, 16 bits resolution, bandwidth 40 kHz |
| 1 Analog out |
0 tp +4 V, 16 bits resolution |
| 1 Digital input, and 3 Digital outputs |
0 to +5 V |
| I-out, and E-out |
analog monitor for cell current and potential |
| AC-out |
±0.5 V sinewave 10µHz-250kHz with variable attenuation |
| Channel-X, and Channel-Y inputs |
±4 V: to record impedance from peripheral devices |
| |
|
| Environment |
|
| Size |
w x d x h = 3 x 35 x 13 cm |
| Weight |
0.7 kg |
| PC requirements |
Windows XP/Vista/7, with free USB port |
| |
|
| Specifications: Ivium-n-Stat main frame |
| Slot positions |
8 (can mount up to 8 s-Modules) |
| Common connectors |
GND and combined emergency off control monitor |
| Power requirements |
20A-frame: 100-240 V, 47-63 Hz, 300 W |
| |
40A-frame: 100-240V, 47-63Hz, 600W |
| Interfacing |
USB 1.1 or 2.0 compliant |
| Size |
w x d x h = 47 x 36 x 14 cm |
| Weight |
6.2 kg (no modules) |
| |
11.8 kg (with 8 modules) |