Skip to Content

IviumStat.XR: Hardware specifications

Printer-friendly versionPrinter-friendly version
Productinfo Hardware specs XR Hardware specs
System Performance: IviumStat XR
Current compliance ±5 A
Maximum output Voltage ± 20V
Potentiostat Bandwidth 1 MHz for small signals, 300kHz for large signals
Stability settings High Speed, Standard, and High Stability
Programmable response filter 1MHz , 100kHz , 10kHz , 1kHz , 10Hz
Signal acquisition dual channel 16 bit ADC, 100.000 samples/sec
   
Potentiostat:  
Applied potential range ±10 V, with 0.33mV resolution
±20 V, with 0.66mV resolution
Applied potential accuracy 0.2%, or 2mV
Current ranges ±10nA to ±10A in 10 steps
High sensitivity current ranges ±1pA, ±10pA, ±100pA, ±1nA
Measured current resolution 0.015% of current range, minimum 0.15fA
   
Galvanostat  
Applied current resolution 0.033% of applied current range
Applied current accuracy 0.2%
Potential ranges ±1mV, ±10mV, ±100mV, ±1V, ±10V
Measured potential resolution 0.003% of potential range, minimum 40nV
   
Impedance Analyser  
Frequency range 10µHz to 2MHz
Amplitude 0.015mV to 1.0V, or 0.03% to 100% of current range
DC offset 16 bit dc offset subtraction, and 2 dc-decoupling filters
Dynamic range 12nV to 10V, and 0.05fA to 5A
   
Electrometer  
Input impedance >1000 Gohm //<8pF
Input bias current <10 pA
Bandwidth >16 MHz
   
Special functions  
Ohmic drop compensation 2 V/current range, 16 bits resolution
Safety features automatic disconnect on internal/external exceptions
   
Peripheral connections  
8 Analog in, and 2 analog out 0 to +4V, 16 bits resolution
2 Digital inputs, and 3 Digital outputs 0 to +5V
I-out, and E-out analog monitor for cell current and potential
AC-out ±0.5V sinewave 10µH-2MHz with variable attenuation
Channel-X, and Channel-Y inputs ±4V: to record impedance from peripheral devices
   
Environment  
Power requirements 100-240V, 47-63Hz, 150VA
Interfacing USB 1.1 or 2.0 compliant
Size w x d x h = 26 x 33 x 12 cm
Weight  4.0 kg
PC requirements Windows XP or Vista, with free USB port

Specifications subject to change, Ivium Technologies © 2009